A survey of technology and temperature control on a neonatal unit in Kathmandu, Nepal

  • R. B. Johanson
  • , D. S. Malla
  • , C. Tuladhar
  • , M. Amatya
  • , Stephen Spencer
  • , P. Rolfe

Research output: Contribution to journalArticlepeer-review

19 Citations (Scopus)

Abstract

An assessment of the incubators in use at the Kathmandu Maternity Hospital neonatal unit was undertaken; this was followed by a prospective survey of neonatal temperatures on the unit. In the incubator assessment 11 studies were carried out in five incubators. Three of the thermostats in the five incubators did not work at all and those in the other two incubators were more than 3°C inaccurate. All the incubator thermometers gave recordings less than the actual temperature (with a range of error: 1.3-4.4°C). Six out of the 11 babies studied were hypothermic (defined as rectal temperature <36°C). In the prospective survey of temperatures a high incidence of hypothermia was found on the neonatal unit at the time of first temperature measurement (64 per cent had a rectal temperature of <36°C). A significant association between admission hypothermia and mortality was noted. Sixteen per cent (10/64) of babies admitted with temperatures <36°C died within the first week, compared to 0 (0/36) of those admitted with temperatures ≥36°C.
Original languageEnglish
Pages (from-to)4-10
Number of pages7
JournalJournal of Tropical Pediatrics
Volume39
Issue number1
DOIs
Publication statusPublished - 1 Feb 1993
Externally publishedYes

Fingerprint

Dive into the research topics of 'A survey of technology and temperature control on a neonatal unit in Kathmandu, Nepal'. Together they form a unique fingerprint.

Cite this